20
Operation
Calling the Calibration Menu
The calibration menu can only be accessed by entering a password.
Password on delivery: »00
The default password on delivery can be changed using the PC software
Dräger CC Vision.
In measuring mode, press the key for at least 4 seconds.
The function for entering the password is selected.
The » « special symbol (for the "enter password" function) is displayed.
Calibration menu functions
Press the key to cancel the active function.
If no key is pressed for 10 minutes, the device automatically returns to measuring
mode.
The display shows »00, with the
first digit flashing.
Use the key to set the flashing digit.
Press the key, the second digit
starts flashing.
Use the key to set the flashing digit.
Press the key, the third digit starts
flashing.
Use the key to set the flashing digit.
Press the key to confirm the
password once it has been set
completely.
The calibration menu functions can
now be selected by pressing the
key.
Press the key to call the selected
function.
Fresh air calibration, refer to page 46
1-button calibration, refer to page 47
Single gas calibration, refer to page 50
02323999_01.eps
OK
OK
OK
OK