Performing a measurement with the Dräger X-am Pump
Required accessories (siehe “Accessories” auf Seite 66):
Dräger X-am Pump
Sampling hose and probes
Pump symbols:
Commissioning and performing the measurement:
See instructions for use of the Dräger X-am Pump.
Viewing pump information:
Open the quick menu (siehe “Calling the Quick Menu” auf Seite 18).
Select and confirm with the button.
The following pump information will be displayed:
— serial number
— pump runtime (current operation)
— pump battery charge
Press the button to return to measuring mode.
Activating or deactivating the pump:
Open the quick menu (siehe “Calling the Quick Menu” auf Seite 18).
Select or and activate or deactivate the pump by pressing the button.
Press the button to return to measuring mode.
Pump battery 100 % charged Warning for pump
(Gas detector can no longer detect
pump.)
Remaining charge of pump
battery: 2/3
Leak test:
Block suction inlet
Remaining charge of pump
battery: 1/3
Leak test:
Release suction inlet
Pump battery discharged
WARNING
No measurement!
If the pump is connected but deactivated, the gas detector is not ready to measure.
The red LEDs on the gas detector flash.
OK