Faults, Cause and Remedy
301 The zero point calibration of
DrägerSensor XXS EC1 is not
valid
Carry out fresh air calibration,
page 46.
302 The span calibration of
DrägerSensor XXS EC1 is not
valid
Carry out span calibration. Carry
out page 50 or fresh air calibra-
tion, page 46.
303 The measured value of
DrägerSensor XXS EC 1 is in
the negative range
Carry out fresh air calibration,
page 46.
304 DrägerSensor XXS EC1 is not
inserted or defective
Check DrägerSensor XXS EC1,
page 59.
305 Error during the function test
with gas (bump test) of
DrägerSensor XXS EC1
Repeat function test. Calibrate
or replace DrägerSensor XXS
EC1, if necessary page 59.
307 Failed rise time test. Repeat rise time test with X-
dock.
308 User parameters not feasible Check configuration of user
parameters and adjust
324 Device incorrectly configured
by Dräger CC-Vision.
Change sensor for applicable
channel with Dräger CC-Vision.
326 Error during warm-up acceler-
ation Dräger Sensor XXS
EC1
Disconnect and reconnect
power pack or replace the sen-
sor. Sensor must not be loaded
with gas within the first 5 min-
utes.
401 The zero point calibration of
DrägerSensor XXS EC2 is not
valid
Carry out fresh air calibration,
page 46.
402 The span calibration of
DrägerSensor XXS EC2 is not
valid
Carry out span calibration,
page 50.
403 The measured value of
DrägerSensor XXS EC 2 is in
the negative range
Carry out fresh air calibration,
page 46.
404 DrägerSensor XXS EC2 is not
inserted or defective
Check DrägerSensor XXS EC2,
page 59.
405 Error during the function test
with gas (bump test) of
DrägerSensor XXS EC2
Repeat function test. Calibrate
or replace DrägerSensor XXS
EC2, if necessary page 59.
406 Failed filter test. Repeat filter test with X-dock.
407 Failed rise time test. Repeat rise time test with X-
dock.
Special symbol »«
and displayed numerical
code:
Cause Remedy