10
Standard instrument configuration
XXS Odorant [ppm] 0 to 40 10 Yes No 20 No Yes
XXS Amine [ppm] 0 to 100 10 Yes No 20 No Yes
XXS COCI
2
[ppm] 0 to 10 0,1 Yes No 0,2 No Yes
XXS O
3
[ppm] 0 to 10 0,1 Yes No 0,2 Yes No
XXS NO
2
LC [ppm] 0 to 50 0,5 Yes No 1,0 Yes No
1) Different settings can be selected to meet customer requirements on delivery. The current setting can
be checked and changed with the Dräger CC Vision software. A version of Dräger CC Vision which can
be used for Dräger X-am 5600 is supplied with the instrument on CD.
2) In the case of O
2
A1 is the lower alarm threshold: an alarm is triggered if the value is too low.
Dräger X-am 5600
1)
1) X-am
®
is a registered trademark of Dräger.
Bump test mode
2)
2) Different settings can be selected by the customer on delivery. The current setting can be
checked and changed with the Dräger CC Vision software.
Extended bump test
Fresh air calibration
2)
3)
3) The fresh air calibration/zero-point calibration is not supported by the DrägerSensor DUAL IR
CO
2
, DrägerSensor IR CO
2
or the DrägerSensor XXS O
3
.
On
Hydrogen offset
4)
4) With activated DrägerSensor XXS H
2
HC (68 12 025) and activated Ex-channel of the
DrägerSensor DUAL IR Ex/CO
2
(68 11 960) or DrägerSensor IR Ex (68 12 180).
On
Operating signal
2) 5)
5) A periodic short flashing indicates the operating capacity of the instrument. If there is no
operating signal, correct operation cannot be guaranteed.
On
Capture range On
Switch off
2)
blocked at A2
LEL factor
2)
ch
4
H
2
4.4 vol. % (4.4 vol. % corresponds to 100 %LEL)
4.0 vol. % (4.0 vol. % corresponds to 100 %LEL)
STEL
2) 6)
7)
(short-term average)
6) STEL: average value of an exposure over a short period, generally 15 minutes.
7) Interpretation only if the sensor is designed for this.
STEL function - disabled
Average value duration = 15 minutes
TWA
2) 7) 8)
(shift average)
8) TWA: shift averages are workplace limit values for generally eight hours per day of exposure for
five days a week during a working life.
TWA function - disabled
Average value duration = 8 hours
Alarm A1
9)
9) Latching and acknowledgement of alarms A1 and A2 can be configured with the Dräger CC
Vision PC software.
can be acknowledged, non-latching, pre-alarm,
rising flank
Alarm A1 at O
2
sensor
9)
cannot be acknowledged, latching, like main alarm,
falling flank
Alarm A2
9)
cannot be acknowledged, latching, main alarm,
rising flank
DrägerSensor
Measuring
range
1)
Alarm A1
1)
Alarm A2
1)
threshold
can be
acknowledged
self-latching
threshold
can be
acknowledged
self-latching