45
The fault message » « appears
and » « is displayed instead of
the measured value on the faulty
measuring channel.
In this case, repeat the function test
with gas or calibrate/adjust the
instrument, page 47.
The function test with gas can also be
carried out automatically.
The "Bump Test Station" is required for
this function, refer to page 45.
Automatic implementation with the Bump Test Station
Prerequisite:
The instrument must first be configured for the automatic function test with gas
(bump test) using the PC software Dräger CC Vision.
Activating the instrument for the automatic function test.
Adjust the test gas concentration (mixed gas) with the Dräger CC-Vision PC
software if it deviates from the following default values – standard on delivery:
50 ppm CO, 15 ppm H
2
S, 2.5 vol. % CH
4
, 18 vol. % O
2
Define which measuring channels should participate in the automatic function test.
All measuring channels participate in the function test by default.
Prepare the Bump Test Station according to the instructions.
Switch on the instrument and insert it into the receptacle of the Bump Test Station
until it engages.
The function test with gas is started automatically. The special symbol » «
(for bump test) flashes.
02633072_01_en.eps
CO2
Vol%
O
2
Vol%
ch4
%LEL
CO
ppm
H
2S
ppm
NH
3
ppm
03833072_01.eps
0
CO2
Vol%
O
2
Vol%
ch4
%UEG
CO
ppm
H
2S
ppm
NH
3
ppm