Replacing the batteries / rechargeable batteries
Switching off the instrument:
Press and hold the key and the
key at the same time.
1 Loosen the screw (2.0 mm hexagon
socket) on the power pack and
remove the power pack.
2a For the battery holder
(order no. 83 22 237):
Replace alkaline batteries or NiMH
rechargeable batteries. Ensure
correct polarity.
2b Completely replace the power pack T4/T4 HC (with sealed rechargeable batteries,
order no. 83 18 704/83 22 244).
Insert the power pack into the instrument and tighten the screw, the instrument
switches on automatically.
After replacing the power pack T4/T4 HC, it is recommended that a complete charging
is carried out.
WARNING
Explosion hazard! To reduce the risk of flammable or explosive atmospheres igniting,
it is essential that the warning notices below are observed:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in areas at risk of an explosion hazard.
Do not mix new batteries with used batteries, and do not mix batteries from different
manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval. Only the following types
may be used:
— Alkaline batteries – T3 – (not rechargeable!)
Panasonic LR6 Powerline
Varta Type 4106
1)
(power one) oder
Varta Type 4006
1)
(industrial)
— Alkaline batteries – T4 – (not rechargeable!)
Duracell Procell MN1500
1)
, Duracell Plus Power MN1500
1)
— NiMH rechargeable batteries – T3 – (rechargeable)
GP 180AAHC
1)
(1800) max. 40 °C ambient temperature.
1) Not subject to BVS10 ATEX E 080X and PFG 10 G 001X performance approval.
WARNING
Danger of explosion!
The Dräger X-am 5600 must only be
operated with battery holder ABT 0100
(X-am 5600), identified with a silver
sticker.