14
During operation
During operation, the measured values for every measured gas are displayed.
In the event of an alarm, the corresponding displays, including the visual, audible
and vibration alarms, are activated. See section "Identifying Alarms"
If a measuring range is exceeded or not reached, the following displays are shown
instead of the measured value display:
After the measuring range of the TOX measuring channels has been exceeded
temporarily (up to one hour), checking the measuring channels is not necessary.
Following an extreme impact load, the display for the Ex- and the CO
2
measuring
range of the IR Sensor must be checked and adjusted with zero gas and span gas
if necessary.
»« (measuring range exceeded) or
»« (measuring range not reached).
NOTICE
Special states in which there is no measuring operation (quick menu, calibration
menu, warm-up of sensors, password input) are indicated by a visual signal (slow
flashing of the alarm LED ).
WARNING
When using an IR Sensor in the Dräger X-am 5600, the zero point and sensitivity
must be adjusted after an impact load that causes a display other than zero for the
IR sensor when the instrument is at fresh air.