Calling the Quick Menu
— Only the fresh air calibration/zero-point calibration is activated in the quick menu on
delivery. The PC software Dräger CC Vision can be used to activate the bump test
for the quick menu and/or the function for displaying and deleting peak values.
In measuring mode, press the key three times.
If no functions have been activated in the quick menu, the instrument remains in
measuring mode.
— You can select the activated functions of the quick menu by pressing the key.
Press the key to call the selected function.
Possible functions of the quick menu
Press the key to cancel the active function and to switch to measuring mode.
— If no key is pressed for 60 seconds, the instrument returns automatically to
measuring mode.
______________________________________________
1) Fresh air calibration / zero-point calibration is not supported by the DrägerSensor DUAL IR CO
2
,
the DrägerSensor IR CO
2
or the DrägerSensor XXS O
3
. A zero-point calibration / adjustment of
these sensors can be conducted using the Dräger CC-Vision PC software. To do so, a suitable
zero gas that is free of carbondioxide and ozone (e.g. N
2
) should be used.
Function test with gas (bump test), refer to page 41.
Fresh air calibration, refer to page 48.
1)
Delete peak values, refer to page 17.
Display pump information, refer to page 21.
Activate or deactivate pump, refer to page 21.
OK