EXT 1824G User’s Manual 8. Troubleshooting
EXT 1824G-UM-COM-EN-03 61
8.3. Battery Pack or Installation Part of Battery is Getting Hot
Symptom
- Battery pack or compartment for installation of battery pack is getting hot.
Possible Causes
- Battery pack failure
- Detector Failure
Solutions
1) Do not use the battery pack
2) Consult with service engineers of DRTECH