EXT 1824G User’s Manual 4. Detector Component
EXT 1824G-UM-COM-EN-03 42
5.2.3. Image Data Retransmission
EXT 1824G can save the image data as a file when detector is disconnected from AP during image data transmission.
User can download the image file or receive the lastest shotted image data by using acquisition mode of Ecali1 after
reconnection.
Figure 5.2. Flow Chart
If user does not use acquisition mode after reconnection, image data is saved as file. Image file cannot be
retransmitted automatically even if acquisition mode is activated.