EasyManua.ls Logo

Dynatronics dynatron solaris 705 plus - Device Specific Procedure

Default Icon
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FINAL QUALITY CHECK (QC) CHECKOFF SHEET
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017 147
Device Specific Procedure
QC, Assembly, Calibration, and Test
D719 Check Sheet -
1 of 9- DS0408A – REV 1
REVISION HISTORY
REV.
CO
DESCRIPTION
APPROVED
DATE
0
12-0726
INITIAL RELEASE
I. BROWN
2012-08-07
1
13-0402
UPDATED MICROCURRENT, 3WL QC CHANGES AND GENERAL UPDATE
DYNATRON SOLARIS
PLUS
D719
FINAL QUALITY CHECK (QC) CHECKOFF SHEET
Repair____RS____Production____ Software Version__________ (initials or mark)
Refer to Test equipment operation manual and unit operating instructions/ manuals
REPAIR/RS
PROD
1
Verify all keys function, all keys have feedback
Verify output turns off when STOP is pressed
Verify output will time out
2
INTERFERENTIAL: 42-55 Vpp (
500ohm 10W load intensity 30
)
a) Static- High/Low (all channels) ± 4 Vpp
b) Target/ Sweep (all channels)
c) Modify Features (mode, time, function, cancel)
d) Verify current limit 50mA maximum with open load
e) Verify conductance
3
PREMODULATED 42-55 Vpp (
500ohm 10W load intensity 30)
a) High/Low (all channels)
b) High/Low alternating (all channels)
c) Modify Features (mode, time, cancel)
d) Verify current limit 50mA maximum with open load
e) Verify conductance
4
RUSSIAN STIM: 42-55 Vpp (
500ohm 10W load intensity 30)
a) One Channel Normal (all channels)
b) Two Channel Reciprocal (channel pairs)
c) Two Channel Co-contraction (channel pairs)
5
BI-PHASIC: 42-55 Vpp (
500ohm 10W load intensity 30)
a) One Channel Normal (all channels)
b) Two Channel Reciprocal (channel pairs)
c) Two Channel Co-contraction (channel pairs)
6
MICROCURRENT PADS:
a) Verify no load, Freq .3Hz, turn off conductance; select Conductance and then
select OFF. At 50 uA +/-4% and at 990 uA +/-1%
b) Verify the polarity changes.
c) Verify conductance.
7
HIGH VOLT PADS:
a) No load, set device to 50 V, verify 42-52 V output.
b) Verify that polarity changes.
8
MULTI-STIM PROBE:
a) No load, Verify DC output, change Pulse Duration to 500. 0-20 mA
b) Verify DC polarity change
c) Verify Hi-Volt output
d) Verify Hi-Volt polarity change
e) Verify micro-current output
f) Verify micro-current polarity change
g) Verify micro-current audible tone
h) Verify UP/DOWN key works
i) Verify ON/OFF key works
j) Verify conductance
Serial Number

Table of Contents

Related product manuals