Safety
94 KOSMOS User Guide
TABLE 9-2 Combined acoustic output reporting table: Reportable mode 2
B-mode (Cardiac, BMI2, 12 cm Depth)
Index label MI
TIS TIB
At
surface
Below
surface
Atsurfa
ce
Below
surface
Maximum index value
0.74 0.50 0.50
Index component value
1: 0.25
2: 0.25
1: 0.25
2: 0.25
1: 0.25
2: 0.25
1: 0.25
2: 0.25
Acoustic parameters
p
r,α
at z
MI
(MPa) 2: 1.06
P (mW) 1: 30.68
2: 30.23
1: 30.68
2: 30.23
P
1x1
(mW) 1: 25.49
2: 25.10
1: 25.49
2: 25.10
z
s
(cm) 1: 4.13
2: 4.20
z
b
(cm) 1: 4.13
2: 4.20
z
MI
(cm) 2: 4.20
z
pii,α
(cm) 2: 4.20
f
awf
(MHz) 2: 2.07 1: 2.07
2: 2.07
1: 2.07
2: 2.07
Other information
prr (Hz) 2: 1589
srr (Hz) 2: 28
n
pps
2: 1
I
pa,α
at z
pii,α
(W/cm
2
)
2: 54.17
I
spta,α
at z
pii,α
or z
sii,α
(mW/cm
2
)
17.24
I
spta
at z
pii
or z
sii
(mW/cm
2
)
31.29
p
r
at z
pii
(MPa) 2: 1.43
Operating
controls
Component 1: UTP73
Component 2: UTP74
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for “at surface” and “below surface” both in the columns related to TIS or
TIB.
NOTE 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS or TIB.
NOTE 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column
related to MI.
NOTE 5 Unshaded cells should have a numerical value. The equipment setting related to the index has
to be entered in the operating control section.
NOTE 6 The depths
z
pii
and z
pii,α
apply to NON-SCANNING MODES, while the depths z
sii
and z
sii,α
apply
to SCANNING MODES.