KOSMOS User Guide 95
Biological safety
TABLE 9-3 Acoustic output reporting table: Reportable mode 3
M-mode (Cardiac, BMI2, 12 cm Depth)
Index label MI
TIS TIB
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.43 5.32E-02 0.11
Index component value
5.32E-02 2.15E-02 5.32E-02 0.11
Acoustic parameters
p
r,α
at z
MI
(MPa) 0.70
P (mW) 4.55 4.55
P
1x1
(mW) 4.11 4.11
z
s
(cm) 5.37
z
b
(cm) 4.80
z
MI
(cm) 5.37
z
pii,α
(cm) 5.37
f
awf
(MHz) 2.72 2.72 2.68
Other information
prr (Hz) 800
srr (Hz) N/A
n
pps
1
I
pa,α
at z
pii,α
(W/cm
2
)
52.08
I
spta,α
at z
pii,α
or z
sii,α
(mW/cm
2
)
16.71
I
spta
at z
pii
or z
sii
(mW/cm
2
)
31.29
p
r
at z
pii
(MPa) 45.72
Operating
controls
UTP4
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for “at surface” and “below surface” both in the columns related to TIS or
TIB.
NOTE 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS or TIB.
NOTE 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column
related to MI.
NOTE 5 Unshaded cells should have a numerical value. The equipment setting related to the index has
to be entered in the operating control section.
NOTE 6 The depths
z
pii
and z
pii,α
apply to NON-SCANNING MODES, while the depths z
sii
and z
sii,α
apply
to SCANNING MODES.