TABLE 9-4 Acoustic output reporting table: Reportable mode 4
M-mode (Cardiac, BMI2, 14 cm Depth)
Index label MI
TIS TIB
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.39 5.33E-02 9.70E-02
Index component value
5.33E-02 2.12E-02 5.33E-02 9.70E-02
Acoustic parameters
p
r,α
at z
MI
(MPa) 0.63
P (mW) 4.60 4.60
P
1x1
(mW) 4.14 4.14
z
s
(cm) 5.50
z
b
(cm) 4.97
z
MI
(cm) 5.50
z
pii,α
(cm) 5.50
f
awf
(MHz) 2.70 2.70 2.67
Other information
prr (Hz) 800
srr (Hz) N/A
n
pps
1
I
pa,α
at z
pii,α
(W/cm
2
)
41.86
I
spta,α
at z
pii,α
or z
sii,α
(mW/cm
2
)
13.64
I
spta
at z
pii
or z
sii
(mW/cm
2
)
38.22
p
r
at z
pii
(MPa) 1.06
Operating
controls
UTP5
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for “at surface” and “below surface” both in the columns related to TIS or
TIB.
NOTE 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS or TIB.
NOTE 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column
related to MI.
NOTE 5 Unshaded cells should have a numerical value. The equipment setting related to the index has
to be entered in the operating control section.
NOTE 6 The depths
z
pii
and z
pii,α
apply to NON-SCANNING MODES, while the depths z
sii
and z
sii,α
apply
to SCANNING MODES.