KOSMOS User Guide 97
Biological safety
TABLE 9-5 Combined acoustic output reporting table: Reportable mode 5
B+C-mode (Abdominal, BMI1, 12 cm depth, smallest color ROI at top)
Index label MI
TIS TIB
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
1.09 1.31 1.31
Index component value
1: 5.47E-02
2: 1.25
1: 5.47E-02
2: 1.25
1: 5.47E-02
2: 1.25
1: 5.47E-02
2: 1.25
Acoustic parameters
p
r,α
at z
MI
(MPa) 2: 1.56
P (mW) 1: 4.68
2: 144.87
1: 4.68
2: 144.87
P
1x1
(mW) 1: 4.23
2: 128.26
1: 4.23
2: 128.26
z
s
(cm) 1: 5.37
2: 2.30
z
b
(cm) 1: 4.80
2: 2.30
z
MI
(cm) 2: 1.97
z
pii,α
(cm) 2: 2.30
f
awf
(MHz) 2: 2.04 1: 2.72
2: 2.05
1: 2.72
2: 2.05
Other information
prr (Hz) 2: 4882
srr (Hz) 2: 28
n
pps
2: 16
I
pa,α
at z
pii,α
(W/cm
2
)
2: 82.58
I
spta,α
at z
pii,α
or z
sii,α
(mW/
cm
2
)
108.55
I
spta
at z
pii
or z
sii
(mW/cm
2
)
152.01
p
r
at z
pii
(MPa) 2: 1.59
Operating
controls
Component 1: UTP4
Component 2: UTP55
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for “at surface” and “below surface both in the columns related to TIS or TIB.
NOTE 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns related to TIS
or TIB.
NOTE 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column related to MI.
NOTE 5 Unshaded cells should have a numerical value. The equipment setting related to the index has to be
entered in the operating control section.
NOTE 6 The depths
z
pii
and z
pii,α
apply to NON-SCANNING MODES, while the depths z
sii
and z
sii,α
apply to
SCANNING MODES.