EasyManua.ls Logo

EchoNous KOSMOS - Page 85

EchoNous KOSMOS
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
KOSMOS Maintenance
78 KOSMOS User Guide
3. At point of use, wipe the cables with an approved presaturated wipe from the
list in Presaturated wipes to ensure effective cleaning.
4. Immerse the ECG clips and leadwires in a cleaning solution from the list in
Cleaning detergent solution for Kosmos ECG patient cable, and soak for at
least 10 minutes. Refer to Cleaning detergent solution for Kosmos ECG
patient cable for the solution concentration and contact time.
5. Place the cable with the solution in a ultrasonicator for at least 10 minutes.
6. Post sonicating, using a standard cleaning brush, vigorously brush all
surfaces of the ECG clips while immersed in the Enzol solution until visibly
clean.
7. Actuate any movable parts while immersed. In addition, flush crevices using a
slip tip syringe filled with prepared cleaning detergent.
8. Remove the ECG clips from the Enzol solution, and run them under running
water for 1 minute. Ensure no gel or any particulate matter is visible after this
cleaning step.
9. Before continuing with disinfection, ensure the Kosmos ECG patient cable is
visibly dry.
TABLE 8-3. Cleaning detergent solution for Kosmos ECG patient cable
Product Company Active Ingredients
Contact
Condition
Enzol Advanced
Sterilization
Products
Borax decahydrate >=5 - <10
Subtilisin >=1 - <5
2 oz. per gallon
solution
20 minutes
immersion
Leadwires Clips

Table of Contents

Other manuals for EchoNous KOSMOS

Related product manuals