KOSMOS User Guide 79
Cleaning and disinfecting
Disinfecting the Kosmos ECG patient cable
Use the following steps to disinfect the Kosmos ECG patient cable. Before
performing the following steps, read the following warnings and cautions.
To disinfect the Kosmos ECG patient cable:
1. After cleaning, choose an low-level disinfectant from the list in Presaturated
wipes, and follow the instructions on the disinfectant label for the minimum
wet contact time.
2. With a new wipe, disinfect the Kosmos ECG patient cable, starting from the
connector end to the clips.
3. Observe the required wet contact time. Monitor the Kosmos ECG patient
cable for wet appearance.
4. Use at least three wipes to ensure effective disinfection.
5. Examine the cable for damage, such as insulation wearing or discoloration. If
damage is evident, discontinue using the Kosmos ECG patient cable.
6. Before reusing the cable, ensure the cable is visibly dry.
Binaural Headset
The following cleaning and disinfection method must be followed for Binaural
Headset:
Always disconnect the USB cable from KOSMOS Torso before cleaning
and disinfecting.
Always use protective eye wear and gloves when disinfecting any
equipment.
Before disinfecting, clean the Kosmos ECG patient cable by following the
appropriate instructions to remove all gels, fluids, and particulates that
may interfere with the disinfection process.
Ensure cable insulation is intact before and after disinfection.
Use only EchoNous-recommended disinfectants. Using a non-
recommended disinfecting wipe can damage the Kosmos ECG patient
cable.