7.1 Initial start-up
In the following cases, the LMI03 starts up with the
No Setup’ alarm screen:
n Turning on a new unit
n After installing a new control PCB in a unit
n After resetting to factory settings
n After replacing the CMOS battery
Carry out initial commissioning
Requirements:
n LMI03 is fully installed.
1. Switch on the unit.
ð
The ‘No Setup
screen appears.
2.
Press to operate the device.
3.
Enter the access code with and confirm with .
ð
The ‘Options
screen (12) appears.
4.
Perform device set-up.
Ä
Chapter 7.2 ‘Set-up’ on page 44
5.
Make the device settings.
Ä
Chapter 7.3 ‘Settings’ on page 49
Installation and start-up
43 MAN049432 Rev. 03-03.2024