7.3 Settings
7.3.1
Conductivity
Calibration of conductivity and temperature
The setting compares the measured conductivity value (actual value) with a preset value
(nominal value). If the actual value is less than the nominal value set for conductivity
, a
metering pump starts up in order to increase the concentration of the solution.
Conductivity measuring cell
Starting point: ‘Options
’ screen (12)
Ä
Chapter 7.1 ‘Initial start-up’ on page 43
1.
Select ‘Settings
’
(13) and confirm with .
2.
Select ‘Conductivity Measuring Cell
’
(40) and confirm with .
3.
Set the calibration using , and .
4.
Confirm the settings with .
ð
Calibration of conductivity and temperature is set.
Additional settings in the Conductivity Settings submenu
n Calibration of conductivity and temperature compensation (41)
Ä
‘Temperature
compensation’ on page 50
Installation and start-up
49 MAN049432 Rev. 03-03.2024