Housing
Data Value Unit
Design:
Wall-mounted unit
Approx. weight:
3 kg
Type of protection (according to DIN 40050):
65 IP
Safety class:
1
Housing feedthroughs (5 threaded unions):
M16x1.5
Measuring cell
Data Value Unit
Oval calotte, streamlined shape with measuring channel
diameter
8 mm
Material: polyether ether ketone
(with BGA and FDA approval)
PEEK
Material: Polypropylene
PP
Dimensions (D x H):
39 x 50 mm
Pressure resistance at 20 °C
(see also pressure/temperature curve):
PN = 10 bar
Temperature resistance, short-term max. (PEEK):
130 °C
Temperature resistance, short-term max. (PP):
90 °C
Chemical resistance:
Resistant to inorganic acids and alkalis, as well as additives in
the range of application concentrations from
0 ... 3 % by weight
Temperature sensor:
NTC resistor in stainless steel thermowell
R25 = 214 kW
Response time of (90% value), approx.
30 s
Material of the temperature sensor in the measuring cell:
1.4571
Sealing element (O-ring):
EPDM
Connector line length:
(Can be extended to max. 20 m using 3 m, 6 m and 20 m
extension cables)
0,2 m
Type of cable:
Special measuring cable with M12 round plug
7 -pole
Measuring cable connection:
- Fixed connection on the sensor side
- Pluggable on the device side with
7-pole M12 round plug
Conductivity measurement
Data Value Unit
Measuring principle: Induction method
Measuring frequency approx.:
10 kHz
Conductivity indicator:
Display with floating decimal point mS/cm indicator
Technical data
75 MAN049432 Rev. 03-03.2024