DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 95 -
Figure 7-2 Maintenance Options
EXCEPT the DICOM Installation, you can select the desired updating option and press Set
to do the updating automatically.
DICOM Install:
You should press Get Serial Number to get the serial number, and then press DICOM Install to
upgrade DICOM.
After installation, press Close and the system pops up prompt information of rebooting the
system.
7.2. Software and Functional Check after the Upgrade
1. Press File on the control panel once to display the file menu.
2. Select Preset, and then select General item to pop up the general preset dialog box.
3. Select System information to see which software/hardware version is installed in the unit.