DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 43 -
3.9.2. Monitor Specifications
LCD: 12.1 inches
Pixel number: 1024*768
Power supply voltage: 3.3 V
Ambient Temperature Relative Humidity Atmospheric Pressure
Operation +5 °C ~ +40 °C 25%~80% 860 hPa-1060 hPa
Transportation -20 °C ~ +55 °C 25%~93% 700 hPa-1060 hPa
Storage -20 °C ~ +55 °C 25%~93% 700 hPa-1060 hPa
Table 3-5 Ambient Conditions
Inverter specifications:
Input voltage: DC12 V
Output voltage: AC1500 V
On control voltage: 5 V
Off control voltage: 0.3 V
WARNING
Dangerous voltage, do not touch the inverter when it’s running.
3.9.3. Software/Option Configuration
Refer to the DUS 60 User Manual, Chapter 5.7.3 General Presetting, for information on
configuring items like hospital name, language, date, time, date format, time format, etc.
For information on configuring user settings, refer to the DUS 60 User Manual, Chapter 5.7.4 –
5.7.8, for presetting information.
3.9.4. User Manual
Check that the correct user manual for the system and software revision is included with the
installation. Specific language versions of the user manual may also be available.