DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 94 -
Chapter 7 Replacement Procedures
WARNING
Service and repairs must only be performed by authorized personnel.
Equipment being returned must be clean and free of blood and other infectious substances.
EDAN maintenance service (EDANMS) state that body fluids must be properly removed from
any part or equipment prior to shipment. EDANMS employees, as well as customers, are
responsible for ensuring that parts/equipment have been properly decontaminated prior to
shipment. Under no circumstance should a part or equipment with visible body fluids be taken or
shipped from a clinic or site (for example, body coils or and ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the
people who will receive or open this package.
NOTE:
The US Department of Transportation (DOT) has ruled that “items what were saturated
and/or dripping with human blood that are now caked with dried blood; or which were
used or intended for use in patient care” are “regulated medical waste” for transportation
purpose and must be transported as hazardous material.
7.1. Ultrasound Application Software Installation Procedure
NOTE: It is possible to update the Ultrasound Application Software via the UPDATE
function.
Tools: U-disk.
Preparations
Before performing the software update/upgrade:
1. Make sure that all system functions are working correct.
2. Check the current software version and installed options.
Software – Installation Procedure (via Maintenance Page):
Connect the U-disk with updating program to the main unit.
Press File and select Preset, and then enter Maintains to display the password window.
Figure 7-1 Password Dialog Box
Enter the password DUS602903G and click OK to display the update window, shown as
below: