DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 62 -
1. Turn the locking handle on the connector housing counterclockwise to the horizontal
position.
2. Firmly grasp the transducer connector and carefully remove it from the system port.
3. Store each transducer in its protective carrying case.
4.5. File Management
Perform the file management as described in section 6.8 of the DUS 60 user manual.
4.6. Software Configuration Checks
Press File and select Preset, and then enter the General Preset to display the general preset
dialog box, this offers different items to check:
Item Task Expected Result(s)
1 General: Check Date and Time setting Date and Time are correct
2
General: Check the Hospital Name is
correct.
Hospital Name is correct
3 General: Check Language settings desired Language is displayed
4 General: Check assignment of Printer desired printer is correct
Press Patient Info to display the patient information dialog box, this offers different items to
check:
Item Task Expected Result(s)
1 User Settings: Check settings of
Name
Patient name is correct
2 User Settings: Check settings of ID Patient ID is correct
3 User Settings: Check all the User
Settings
Settings assigned as desired by the customer
Press File and select Preset, and then enter the every exam preset to display the corresponding
preset dialog box, this offers different items to check:
Item Task Expected Result(s)
1
Application Parameters:
Check all settings for all applications
setting assigned as desired by the customer
Press File and select Preset, and then enter formula preset to display the formula preset dialog
box, this offers different items to check:
Item Task Expected Result(s)
1
Global Parameters:
Check all settings
Settings assigned as desired by the customer
4.7. Peripheral Checks
Check that peripherals work as described below:
Item Task Expected Result(s)
1 Press Print on the report page The report is printed on printer.
2 Press Print on the control panel
The image displayed on the screen is printed on
printer.
3 Check the VGA monitor The monitor displays the layout in normal.
4 Check the video monitor The monitor displays the image in normal.