DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 69 -
Step 11. Take down the caps of the knobs and TGC sliders, covers of the LED, and silicone
keyboard to get the separate standard PC board and keyboard board.
Figure 5-6 Keyboard Board
5.2.2. Battery
Step 1. Turn the flicker counterclockwise;
Step 2. Pull the battery out.
Step 1 Step 2