DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 96 -
Figure 7-3 System Information
4. Restart the system and perform basic functional checks to ensure system is functioning
normally.
7.3. Replacing Fuses
You can replace the fuses if necessary.
Step 1. Pull out the fuse-box using tweezers;
Step 2. Use the tweezers to pull the fuses out through the small hole on the bottom of the box;
Step 3. Put new fuses (φ5×20, T3.15AH250V) provided by EDAN in position, and reposition
the fuse-box.
Step 1