DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 70 -
5.2.3. Power Board
Step 1. Put the system rear side up on a foam to avoid scraping the LCD, and unbolt the screw on
the left side of the system;
Step 1
Step 2. Overturn the system to make the front side up and unbolt the four screws to separate the
main unit from the rear cover;
Step 2
Step 3. Pull the rear cover outwards on the left panel side with one hand, and lift the main unit
with the other;
Step 4. Lift the separate main unit with both hands;
Step 3 Step 4
Screw position
Screw position