DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 71 -
Step 5. Unbolt the screws on the power module cover and disconnect the two cables connected
with the main board;
Step 5
Step 6. Overturn the system to make the rear side up, disconnect the cable and unbolt the four
screws to get the separate power module;
Step 6
Screw position
Disconnect the two cables
Disconnect
the cable
Screw position