DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 13 -
Acoustic Output Reporting Table
Transducer Mode C613UA Operation Mode PW Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.4034 0.447 1.5775 0.5415
P
r.3
MPa 0.9316
W
0
mW 17.62 17.62 17.62
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
Z
1
(cm)
Z
bp
(cm)
Z
sp
(cm) 1.4
Z@PII
.3max
(cm) 1.4
deq(Z
sp
) (cm) 0.0144
f
c
(MHz) 5.3316 5.3316 5.3316 5.3316
X(cm) 0.896 0.896 0.896
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
0.58 0.58 0.58
PD (usec) 1.7398
PRF (Hz) 6361
P
r
@PII
max
(MPa) 1.3869
d
eq
@PII
max
(cm) 0.0144
FL
x
(cm) 0.469 0.469 Focal
Length
Fl
y
(cm) 3.8273 3.8273
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.053
Control
Conditions
Control1
AP=15; Frequency=5.5MHz;
Depth=29mm; Focus=10mm;