DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 15 -
Acoustic Output Reporting Table
Transducer Mode C343UA
Operation Mode B+M Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.9043 0.1146 0.0248 0.0961 0.4581
P
r.3
MPa 1.557
W
0
mW 32.71 3.244 35.954
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
1.75
Z
1
(cm) 2.9389
Z
bp
(cm) 2.9389
Z
sp
(cm) 4.85
Z@PII
.3max
(cm) 4.85
deq(Z
sp
) (cm) 0.236
f
c
(MHz) 2.9652 2.9652 2.9652 2.9652 2.9652
X(cm) 2.061 2.061 2.061 2.061
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
1.5 1.5 1.5 1.5
PD (usec) 0.5188
PRF (Hz) 1694
P
r
@PII
max
(MPa) 2.5988
d
eq
@PII
max
(cm) 0.236
FL
x
(cm) 1.5726 1.5726 1.5726Focal
Length
Fl
y
(cm) 21.5005 21.5005 21.5005
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.1552
Control
Conditions
Control1
AP=15; Frequency=2.0MHz;
Depth=98mm; Focus=60mm;