DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 18 -
Acoustic Output Reporting Table
Transducer Mode L763UA Operation Mode B+M Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
≤1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.8131 0.0653 0.0132 0.0443 0.2448
P
r.3
MPa 1.816
W
0
mW 11.08 1.072 12.152
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
0.565
Z
1
(cm) 1.8587
Z
bp
(cm) 1.8587
Z
sp
(cm) 2.25
Z@PII
.3max
(cm) 2.25
deq(Z
sp
) (cm) 0.5263
f
c
(MHz) 4.9898 4.9898 4.9898 4.9898 4.9898
X(cm) 2.016 2.016 2.016 2.016
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
0.6 0.6 0.6 0.6
PD (usec) 0.3435
PRF (Hz) 2295
P
r
@PII
max
(MPa) 2.6699
d
eq
@PII
max
(cm) 0.5263
FL
x
(cm) 0.3902 0.3902 0.3902Focal
Length
Fl
y
(cm) 2.2322 2.2322 2.2322
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.12
Control
Conditions
Control1
AP=15; Frequency=6.0MHz;
Depth=80mm; Focus=30mm;