DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 33 -
Acoustic Output Reporting Table
Transducer Mode L743UA
Operation Mode B+M Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.7874 0.0866 0.0183 0.0368 0.2517
P
r.3
MPa 1.985
W
0
mW 8.0099 0.72 0.72 8.2799
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
Z
1
(cm)
Z
bp
(cm)
Z
sp
(cm) 2.6
Z@PII
.3max
(cm) 2.6
deq(Z
sp
) (cm) 0.6139
f
c
(MHz) 5.2224 5.2224 5.2224 5.2224 5.2224
X(cm) 1.344 1.344 1.344 1.344
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
0.6 0.6 0.6 0.6
PD (usec) 0.3289
PRF (Hz) 2295
P
r
@PII
max
(MPa) 3.0623
d
eq
@PII
max
(cm) 0.6139
FL
x
(cm) 0.7188 0.7188 0.7188Focal
Length
Fl
y
(cm) 7.8222 7.8222 7.8222
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.1536
Control
Conditions
Control1
AP=15; Frequency=6.0MHz;
Depth=78mm; Focus=35mm;

Table of Contents