DUS 60 Digital Ultrasonic Diagnostic Imaging System Service Manual
- 36 -
Acoustic Output Reporting Table
Transducer Mode E743UA
Operation Mode B+M Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
≤1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.8577 0.0881 0.0183 0.0371 0.2521
P
r.3
MPa 1.967
W
0
mW 8.2969 0.75 0.75 9.0469
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
Z
1
(cm)
Z
bp
(cm)
Z
sp
(cm) 2.9
Z@PII
.3max
(cm) 2.9
deq(Z
sp
) (cm) 0.6139
f
c
(MHz) 5.262 5.262 5.262 5.262 5.262
X(cm) 1.344 1.344 1.344 1.344
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
0.6 0.6 0.6 0.6
PD (usec) 0.3378
PRF (Hz) 2287
P
r
@PII
max
(MPa) 3.005
d
eq
@PII
max
(cm) 0.6132
FL
x
(cm) 0.7188 0.7188 0.7188Focal
Length
Fl
y
(cm) 7.8222 7.8222 7.8222
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.11907
Control
Conditions
Control1
AP=15; Frequency=6.0MHz;
Depth=78mm; Focus=35mm;