iM80 Patient Monitor Service Manual Safety Guidance
- 4 -
Chapter 2 Safety Guidance
2.1 Introduction
This service manual is a reference for periodic preventive maintenance and corrective service
procedures for the iM80 patient monitor. It provides information on troubleshooting, assembly
procedures, and instructions for functional testing as well as performance verification. The
manual is intended for use only by technically qualified service personnel.
WARNING
Please follow the instructions exactly in accordance with this manual during service.
Failure to do so might result in damage to the monitor or personal injury.
2.2 General Information
iM80 Patient Monitor (hereinafter called monitor) is designed in accordance with the
international safety requirements in IEC/ EN 60601-1 for medical electrical equipment.
Classification information of this equipment is as follows:
Anti-electroshock Type Class I equipment and internal powered equipment
Anti-electroshock Degree NIBP, SpO
2
, CO
2
, AG BF
ECG, TEMP, IBP, C.O. CF
Ingress Protection IPX1
Degree of Safety in Presence of
Flammable Gases
Not suitable for use in presence of flammable gases
Working System Continuous operation equipment
EMC Type Group 1 Class A
2.3 Safety Precautions
To avoid possible injury, please observe the following precautions during the operation of the
monitor.