iM80 Patient Monitor Service Manual Disassembling the Monitor
- 47 -
To disassemble the CO
2
module:
Assemble the CO
2
module in the reverse order.
8.5.5 Replacing NIBP Module
To disassemble the NIBP module:
Assemble the NIBP module in the reverse order.
Separate module bracket assembly
as described in section 8.3
Unscrew the screws
securing NIBP module
Disconnect
wires
connecting NIBP module
Separate the front and back
assemblies
Unscrew the screws
securing CO
2
module
Disconnect
wires
connecting CO
2
module
Wire connecting main board
Wire connecting main board
Wire connecting
main board