iM80 Patient Monitor Service Manual Test and Maintenance
- 20 -
Applied
Part
Normal
Condition
Single Fault
Condition
Patient Leakage Current
CF
AC: <0.01 mA
DC: <0.01 mA
AC: <0.05 mA
DC: <0.05 mA
BF
AC: <0.1 mA
DC: <0.01 mA
AC: <0.5 mA
DC: <0.05 mA
Patient Leakage Current (Mains on
Applied Parts)
CF <0.05 mA
BF <5 mA
Patient Auxiliary Current
CF
AC: <0.01 mA
DC: <0.01 mA
AC: <0.05 mA
DC: <0.05 mA
BF
AC: <0.1 mA
DC: <0.01 mA
AC: <0.5 mA
DC: <0.05 mA
4.3.5 Patient Leakage Current- Single Fault Condition (S.F.C) Mains on
Applied Part
NOTE:
The following test is based on test with the Fluke 601 pro series safety analyzer. This
device allows applying a 110% mains voltage between the applied part and the device
PE. When testing with other device, you may need to apply the 110% mains voltage
manually.