iM80 Patient Monitor Service Manual Troubleshooting
- 37 -
fluctuation.
the patient; adjust the ruler.
The
IBP communication failure.
Failure in the PCBA of IBP
module or disconnection of
Check the connection of IBP
module and communication board
or change the IBP module.
7.16 AG Monitoring Failures
Phenomenon Possible Cause Solution
No waveform
Power off, and reconnect GAS
module.
GAS module defective. Replace GAS module.
AG waveform is smooth.
GAS module is in Standby
mode.
Measurement mode.
The tube is jammed or not
connected well.
Power off, and clean or replace the
tube.
automatically.
Normal phenomenon.
zeroed for a long time, thus
inaccurate.
Zero the GAS module.
Measure again when the system
enters measurement mode after 10
minutes of warm-up.
7.17 Technological Alarms
For details on technological alarms, please refer to relevant section in the user manual.