EasyManua.ls Logo

EDAN iM80 - Spo Monitoring Failures; NIBP Monitoring Failures; TEMP Monitoring Failures

EDAN iM80
59 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
iM80 Patient Monitor Service Manual Troubleshooting
- 35 -
Frequent patient activity. Quiet the patient.
RESP waveform is
disabled.
Enable the RESP waveform in
system menu.
RESP waveform amplitude
faint.
Adjust waveform amplitude
multiple in RESP menu.
ECG module defective. Replace ECG module.
HR inaccurate, HR
abnormal, and ST analysis
incorrect.
Measurement waveforms
defective.
Adjust the connection.
7.11 SpO
2
Monitoring Failures
Phenomenon Possible Cause Solution
No SpO
2
waveform.
SpO
2
transducer or module
defective.
Replace the SpO
2
transducer.
Strong SpO
2
waveform
interference.
Patient activity. Quiet the patient.
Extremely strong ambient
light.
Weaken ambient light.
SpO
2
value is inaccurate.
The patient is injected with
dye.
Eliminate dye before measurement.
7.12 NIBP Monitoring Failures
Phenomenon Possible Cause Solution
The cuff fails to be inflated.
The
cuff is folded or air
leak occurs.
Repair the cuff.
No NIBP numeric value.
Loose cuff or patient
activity.
Tie up the cuff or quiet the patient.
Big error in NIBP numeric
value.
Cuff size unsuitable for
patient.
Use appropriate cuff.
NIBP module defective. Replace NIBP module.
7.13 TEMP Monitoring Failures
Phenomenon Possible Cause Solution
TEMP numeric value
incorrect.
Bad connection of TEMP
transducer.
Connect and fix the TEMP
transducer well.

Table of Contents

Other manuals for EDAN iM80

Related product manuals