iM80 Patient Monitor Service Manual Troubleshooting
- 35 -
Frequent patient activity. Quiet the patient.
RESP waveform is
disabled.
Enable the RESP waveform in
system menu.
RESP waveform amplitude
faint.
Adjust waveform amplitude
multiple in RESP menu.
ECG module defective. Replace ECG module.
HR inaccurate, HR
abnormal, and ST analysis
incorrect.
Measurement waveforms
defective.
Adjust the connection.
7.11 SpO
2
Monitoring Failures
Phenomenon Possible Cause Solution
No SpO
2
waveform.
SpO
2
transducer or module
defective.
Replace the SpO
2
transducer.
Strong SpO
2
waveform
interference.
Patient activity. Quiet the patient.
light.
Weaken ambient light.
SpO
2
value is inaccurate.
The patient is injected with
dye.
Eliminate dye before measurement.
7.12 NIBP Monitoring Failures
Phenomenon Possible Cause Solution
The cuff fails to be inflated.
The
leak occurs.
Repair the cuff.
No NIBP numeric value.
activity.
Tie up the cuff or quiet the patient.
Big error in NIBP numeric
patient.
Use appropriate cuff.
NIBP module defective. Replace NIBP module.
7.13 TEMP Monitoring Failures
Phenomenon Possible Cause Solution