iM80 Patient Monitor Service Manual Installation
- 7 -
Chapter 3 Installation
WARNING
Only qualified service engineers should install this equipment.
3.1 Environment Requirements
Working
Temperature 5 ºC ~ 40 ºC ( 41ºF ~ 104ºF)
Relative Humidity 25% ~ 80% (non-condensing)
Atmospheric Pressure 860hPa ~ 1060hPa
Storage
Temperature -20 ºC ~ 55 ºC
Relative Humidity 25% ~ 93% (non-condensing)
Atmospheric Pressure 700hPa ~ 1060hPa
NOTE:
1 Do not install the monitor in close proximity to flammable anesthetics.
2 Keep the environment clean and keep the device away from corrosive medicine.
Prevent the device from vibration, high temperature, humidity and exposure to the
sun.
3.2 Electrical Requirements
Operating Voltage 100V-240V ~
Operating Frequency 50Hz/60Hz
Current 1.4A-0.7A