iM80 Patient Monitor Service Manual Test and Maintenance
- 14 -
Cylinder;
Manometer.
Procedure:
1. Access Menu > Maintenance > User Maintain by inputting the password ABC.
2. Connect the equipment as shown below:
Manometer
NIBP module
/ Patient
monitor
Cylinder
T-fitting
NIBP
extension
tube or hose
Tube
Tube
Figure 4-2 Diagram for NIBP Calibration
3. Select NIBP Maintain > Calibrate.
4. Apply fixed static pressure on the monitor with the help of the manometer. Compare the
displayed values on the monitor with the manometer values.
5. A difference within the range of ±3mmHg is reasonable.
4.2.6 TEMP Performance Test
This test checks the performance of the TEMP measurement.
Tools required: resistance box.
Procedure:
1. Switch on the monitor and the resistance box.
2. Set the probe type on the monitor to YSI-10K, and respectively connect the probes to
channel T1 and T2 connectors. And then connect the probes with the resistance box.
3. Set the resistance value to (6017Ω) 37 in the resistance box.
4. A difference within the range of ±0.1 is reasonable.