iM80 Patient Monitor Service Manual Installation
- 8 -
3.3 Safety Requirements
CAUTION
1 SHOCK HAZARDTo protect patients and medical staff, the power receptacle must
be well grounded. Never adapt the three-prong plug from the monitor to fit a
two-socket outlet.
2 Do not simultaneously touch the signal input or output connector and the patient.
3 The monitor and equipment connected to the monitor should be equipotential to
ensure effective grounding.
4 Do not switch on the monitor until all units and accessories have been properly
connected and verified.
3.4 Installing the Monitor
- To install the monitor on a flat surface.
Place the monitor on a flat surface. Make sure the surface does not vibrate, and is free of
corrosive medicine and dust.
iM80 on a Flat Surface
- To install the monitor on a trolley.
If the user wants to install the monitor on a trolley, please refer to the assembling instruction
delivered with the trolley for details.