iM80 Patient Monitor Service Manual Disassembling the Monitor
- 45 -
Assemble the main board in the reverse order.
1 Wire connecting SpO
2
module
2 Wire connecting ECG module
3 Wire connecting CO
2
module
4 Wire connecting power module
5 Wire connecting key board
6 Wire connecting interface board (big)
7 Wire connecting interface board (small)
8 Wire connecting NIBP module
9 Wire connecting interface board (big)
Separate the front and back
assemblies
Unscrew the scr
ews
securing main board
Disconnect
wires on
main board
4
5
3
2
1
6
7
8
9