3. Before Use
3.1 Intended Use
E-connect S is exclusively designed for dentists for use with dental root canal
instruments in continuous rotation and in reciprocating movement with integrated
apex locator.
This device must only be used in hospital environments, clinics or dental offices
by qualified dental personnel.
3.2 Contraindications
The integrated apex locator of the E-connect S is contraindicated in cases
where patient/user carry medical implants such as pace makers or cochlear implants
etc.
Do not use the device for implants or other non-endodontic dental procedures.
Safety and effectiveness have not been established in pregnant women and
children.
Read the following warnings before use:
1. The device must not be placed in humid surroundings or anywhere where it can
come into contact with any type of liquids.
2. Do not expose the device to direct or indirect heat sources. The device must be
operated and stored in a safe environment.
3. The device requires special precautions with regard to electromagnetic
compatibility (EMC) and must be installed and operated in strict compliance with the
EMC information. In particular, do not use the device in the vicinity of fluorescent
lamps, radio transmitters, remote controls, portable or mobile RF communication
devices and do not charge, operate or store at high temperatures. Comply with the
specified operating and storage conditions.
4. Gloves and a rubber dam are compulsory during treatment.
5. If irregularities occur in the device during treatment, switch it off. Contact the
agency.
6. Never open or repair the device yourself,otherwise, void the warranty.