Guidance and manufacturer’s declaration – electromagnetic emissions
The E-connect S is intended for use in the electromagnetic environment specified below.
The customer or the user of the E-connect S should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR
11
Group 1
The E-connect S uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR
11
Class B
The E-connect S is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Not
applicable
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Not
applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The E-connect S is intended for use in the electromagnetic environment specified below.
The customer or the user of the E-connect S should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment - guidance