Page 45 / 59
8.Maintenance
Only the components above can be autoclaved.
Cleaning
Disinfection Lubrication
Packing
Sterilization
Storage
Cleaning: Clean the components with running water with a soft cloth.
Disinfection: Wipe the components with a piece of gauze that has been
dampened with Ethanol for Disinfection (Ethanol 70 to 80 vol%) and wrung out
thoroughly.
Lubrication: Only the contra angle needs to be
lubricated
Before autoclaving, the
contra angle must be
lubricated.
Attaching the spray nozzle to
oil can and contra angle,
press the oil can button more
than 3 seconds, till all the
black oil flow out from the
head of the contra angle.
Packing: Pack the components in“Sterilization pouches”.
Sterilization: Steam sterilization at 134°C at least 4 minutes or
steam sterilization at 121°C at least 35 minutes.