2. In the Field Communicator application, tap HART.
3. Tap Yes when prompted if you want the Trex unit to power the device.
4. Tap Transmitter or Positioner.
If you selected Positioner, the Trex unit provides power and 4 mA.
5. Follow any additional prompts.
You may see additional prompts if the Trex unit cannot connect to the device. For
example, you may be prompted to review the polling address. For a positioner, you
may be asked to increase current (mA) to increase the value from the original 4 mA.
The Online menu or the device dashboard displays, depending on the device type.
Related information
Wiring diagrams for HART devices and the Field Communicator application
Wiring diagrams for the Smart Wireless THUM
Adapter and the Field Communicator
application
Disconnect a device
HART communication terminals
Internal resistors
3.12.1 Power and connect to a Smart Wireless THUM adapter
The Trex unit must have the Device Communicator Plus communication module to power
a THUM adapter.
When performing a bench configuration, the THUM adapter should be attached to a
device. However, the Trex unit can power a THUM adapter when it is not connected to a
HART device.
To connect to a device that has a THUM adapter attached, see Section 3.13 or Section 3.12.
CAUTION!
Remove the USB cable from the Trex unit before connecting to a device.
Procedure
1. Connect the lead set to the HART + pwr terminals on the Trex unit and the red/black
wires on the THUM adapter.