3.15.1 Create a HART configuration
Create a new configuration for a specific device type and revision. You can then apply this
configuration to multiple devices.
Procedure
1. In the Field Communicator application, tap HART Offline.
2. Tap Offline Config > New.
3. Tap a manufacturer.
4. Tap a device type.
5. Tap a device revision.
6. Tap OK.
7. Select and edit the necessary parameters.
8. Tap OK to mark the parameter for writing.
Note
Unmarked parameters are saved but not sent when you transfer the configuration to a
connected device.
9. Tap Save when you finish editing the parameters.
10. Enter a name for the configuration, and tap OK.
The configuration file is saved.
3.15.2 Edit a HART configuration
Prerequisites
Note
Do not edit a configuration on the Trex unit while you are connected to the Configuration
Management PC application.
Procedure
1. In the Field Communicator application, tap HART Offline.
2. Tap Offline Config > Edit Existing.
3. Tap the column heading to sort configurations by File Name or Tag. Select the
configuration you want to edit, and tap OK.
4. Select and edit the parameters.
5. Tap the checkbox to mark the parameter for writing.
Note
Unmarked parameters are saved but not sent when you transfer the configuration to a
connected device.
Field Communicator application
90 User Guide

Table of Contents

Other manuals for Emerson AMS Trex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson AMS Trex and is the answer not in the manual?

Emerson AMS Trex Specifications

General IconGeneral
BrandEmerson
ModelAMS Trex
CategoryCell Phone
LanguageEnglish

Related product manuals