Any 5-digit 14-segment unit may be entered for a custom rate unit (example: mL).
Any 7-digit 14-segment unit may be entered for a custom total or grand total unit
(examples: GALLONS, BOTTLES, DRUMS).
When selected for total or grand total, a custom unit does not allow a multiplier prefix.
Enter a total or grand total conversion factor to define a custom total or grand total unit.
Refer to Custom units total conversion factor (totCF) for details.
Custom units entry (USEr)Figure 5-13:
Alphanumeric values are set by using the Right button to select the digit, the Up and Right
arrow buttons to select the digit reading, and the Enter button to confirm and select the
next digit.
Refer to Section 5.4 for details on setting alphanumeric labels.
Press Menu button to exit this menu without saving changes.
Custom units rate conversion factor (rAtCF)
The rate conversion factor is only used when the Units for rate have been set to custom
(CUST). This menu will not appear if standard display units are selected for the rate unit.
Rate Conversion Factor is used to convert to a custom unit of rate display. For example, to
display rate as quantity of 2.5 gallon containers when the K-Factor units are set to gallons,
enter a conversion factor of 2.500.
Press the Enter button, at any time, to accept a setting or Menu button to exit without
saving changes.
Setup and programming
46 LTM Internal Display