8. Repeat Step 1 through Step 6 for the other transducer(s) being careful to note the
location of each transducer in the meter assembly.
9. Run the field hydrostatic test.
10. Reverse the steps above to reinstall the transducers into their appropriate ports.
Before reinstalling the transducer assemblies, ensure the transducer ports, mounts,
and transducer holders are clean and free of debris. Apply a small amount of Nickel
antiseize compound to the outer threads of the transducer holders before installing
them into the mounts.
1.3 Routine maintenance
Routine maintenance operations requires adherence to all applicable regulations and laws
and safety training for personnel to perform the maintenance operations. Review your
organization’s best practices procedures before performing routine maintenance.
1.3.1 Maintenance logs and reports
To monitor the performance health of the meter, and ensure it is operating within
acceptable specifications, routine diagnostics should be performed. Collecting a
maintenance log gives you a snapshot of the current health of the meter and you can
compare the inspection reports from previously saved logs. Use the Logs/Reports menu
and click Maintenance Logs and Reports. Daniel MeterLink displays the Maintenance Logs
and Reports dialog box. Choose the time duration, log format and collection rate for the
output file and click the Start button. You can open the file immediately after it is
generated or view it at a later time. It is recommended that a Maintenance log be collected
after an upset in the system.
In establishing a baseline to be used for the trending of the meter diagnostics, it is very
helpful if a set of log files are collected immediately after the meter has been installed in
the field. Preferably, collect the log files at several velocities within the operating range of
the meter. This helps establish that the flow profile is relatively constant throughout the
meters operating range (except velocities below 3 ft/sec where the profile may vary).
Maintenance
4 Gas Ultrasonic Flow Meters

Table of Contents

Other manuals for Emerson Daniel 3410 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Daniel 3410 Series and is the answer not in the manual?

Emerson Daniel 3410 Series Specifications

General IconGeneral
BrandEmerson
ModelDaniel 3410 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals