Daniel MeterLink Transducer Swap-out WizardFigure 3-7:
Chord “L” Dimension Calculation
The chord “L” dimension is calculated from the meter housing length as well as the
transducer pair lengths, mount lengths, holder lengths, and stalk lengths as shown in
Equation 3-1. The transducer lengths are etched on the transducers. Likewise, the lengths
of the mounts, stalk assemblies, and transducer holders are also etched on the individual
components. The length of the meter body is found on the original calibration sheet
supplied with the meter.
Chord “L” Dimension Equation 3-1:
ℎ
=
sin
+
1
+
2
1
1
1
2
2
2
where
L
chord
chord “L” dimension (in) (LA ... LD)
L
MeterHousing
meter housing length (in)
L
Mount1
transducer 1 mount length(in)
L
Mount2
transducer 2 mount length (in)
L
Xdcr1
transducer 1 length (in)
L
Xdcr2
transducer 2 length (in)
L
Stalk1
transducer 1 stalk length (in)
L
Stalk2
transducer 2 stalk length (in)
L
Hldr1
transducer 1 holder length (in)
Meter repairs
Maintenance and Troubleshooting manual 55